Hangzhou Huihe Equipment Co. Ltd:2013-11-27
Computer automatic water bath top and bottom tank's autoclave operation all process video
(1) Autoclave's operating process index and maximum working pressure, the highest working temperature.
(2) Autoclave's operating methods and procedures and matters needing attention item.
(3) Should check the project and parts, mainly in the operation of the production and operation of possible anomalies and prevention measures.
(4) Equipment downtime and maintenance methods
(5) No more than design pressure and the maximum temperature operation.
(6) The operator should be after passing the exam of the training, can let him to operate the
sterilization equipment
2. Autoclave's accessories has safety valves, pressure gauge, thermometer, PLC control system, etc. (Note: Only computer automatic autoclave and computer semi-automatic have PLC control system, Want to know about PLC control system relevant data may be click here to download
http://www.jswift.cn/PicUpFile/20135161671930226.zip)
Autoclave's PLC control system
Autoclave's pressure gauge
Autoclave's thermometer
Warm note:
1. Should strengthen the maintenance and calibration during use.
2. Safety valve and pressure jump is equal to the design pressure, it should maintain sensitive, reliable and it should prevent adjusted freely.
3. Pressure gauge, thermometer's precision grade of 1.5, within the tolerance allowed differences is normal.
4. No seal pressure gauges shall not be used, Gauge indicates failure, the scale is not clear, dial is broken, pointer is not back to zero after pressure relief, seal damage, and so on and so forth should be replaced immediately.
5. The thermometer should conduct regular inspection, before use should be in a standard thermometer calibration and it's best nuclear once a year, Mercury column fracture and differ with normal temperature more than 0.5 celsius should be repairing or replacement.
6. At least once every six months the external inspection, every year at least a full inspection, the preparing work before test and inspection items, by "autoclave operating instructions" and the relevant provisions, inspection report filed for the record.
7. During use should be often to bevel self-locking wedge, roller, pan ear, swivel flange, such as sliding and rolling contact parts, daub grease and oil, in order to make sure the friction surface in the lubrication state.
8. If flange sealing ring on the surface of the pot body aging, defect and wear too big, then should be replaced in a timely manner, in order to ensure the seal and prevent leakage.
9. Equipment downtime, sterilization pot and a pot of inside and outside surface should be clean. Leakage processing brush with antirust paint, safety inspection control instrument shall cover sealing protection.
Note: this autoclave's using and maintenance on May 16, 2013 revision, according to the customer usage instructions for the change in recent years nearly, to help users better maintain their own products, so as to extend the life of the product cost savings!